| Standard Parameter |
5‐lead ECG, RESP, HR, NIBP, SpO<sub>2</sub>(F‐5), 1‐TEMP, PR |
| Patient Groups |
Neonate, Pediatric, Adults |
| Trace |
9 waveform (ECG,SpO<sub>2</sub>, RESP) |
| ECG Lead Mode |
5–lead (R, L, F, N, C) for leads I, II, III, V, aVR, aVF, aVL |
| ECG Waveform |
3 and 7 channel selectable |
| ECG Gain selection |
0.5mm/mv </br> 1mm/mv </br> 2mm/mv |
| ECG Sweep Speed |
12.5 mm/s </br> 25 mm/s </br> 50 mm/s |
| Heart Rate range |
Adult : 15 ~ 300 bpm </br> Neonate /Pediatric : 15 bpm ~ 350 bpm </br> Alarm range : 15 ~ 350 bpm |
| ECG Accuracy |
+1 bpm |
| ECG Resolution |
1 bpm |
| Protection |
withstands 4000VAC/50 voltage isolation |
| S‐T segment analysis |
Range : -2.0 mV ~ +2.0 mV </br> Accuracy: ‐0.8mV ~ +0.8mv </br>
Error: +0.02Mv |
| Arrhythmia analysis |
NA |
| Scaling |
1mv, +3% |
| SpO<sub>2</sub> Measurement range |
0% ~ 100% </br> Alarm range : 0% ~ 100% |
| SpO<sub>2</sub> Accuracy |
±2 % (70 % – 100 %) </br> 0 – 69% unspecified |
| SpO<sub>2</sub> Pulse Rate |
Range : 20bpm ~ 300bpm </br> Resolution: 1bpm </br> Error: +1bpm |
| NIBP Method |
Digital Automatic Oscillometric |
| NIBP Operation Mode |
Manual/Automatic/Continuous for Systolic, Diastolic and Mean |
| Auto Measurement |
Adjustable </br> Duration : 1 ~ 480 mins |
| Range Of Systolic Pressure |
Adult: 40 ~ 270mmHg </br> Pediatric: 40 ~ 220mmHg </br> Neonate: 40 ~ 135 mmHg |
| Range Of Diastolic Pressure |
Adult: 10 ~ 215 mmHg </br> Pediatric: 10 ~ 150 mmHg </br> Neonate :10 ~ 100 mmHg |
| Over‐pressure Protection |
Double safety protection |
| Respiratory Method |
Thoracic impedance |
| Respiratory Measurement range |
Adult: 7 ~ 120 rpm </br> Neonate/Pediatric: 7 ~ 150 rpm |
| Apnea Alarm |
10s ~ 40s |
| Respiratory Resolution |
1 rpm |
| Respiratory Accuracy |
±2 bpm |
| Temp Measurement range |
5 ℃ ~ 50 ℃ |
| Average Measuring Time |
< 10 sec |
| Temp Resolution |
0.1℃ |
| Temp Accuracy |
± 0.1℃ |
| Display |
12.1’’color TFT |
| Resolution |
800 x 600 dot pixels |
| Trend |
5 s/piece, 8 hours </br> 5min/piece, 1000 hours |
| Alarm |
User‐adjustable 3‐level Audio-visual alarm. |
| Safety Standard |
Class I, type CF |
| Working Temperature |
0 ~ 40℃ |
| Relative Humidity |
≤ 85 % |
| Power Supply |
AC 100 ‐ 240 V, 50/60 Hz |
| Battery |
Rechargeable Lead acid cell, 12 V/4 Ah |
| Storage |
NIBP : 1000 groups </br> Alarms : 200 groups </br> Holographic waveform : 1 hour |
| Dimension |
280 × 420 × 480 mm |
| Weight |
5 Kgs |